Temperatur der Umgebung.
ambient temperature
The ITS-90 scale applies to temperature data with this refType by convention. Temperatures for which a different scale is applicable must have an additional refType temp_ (see definitions there).
Per Konvention gilt für Temperaturangaben mit diesem refType die ITS-90-Skala. Temperaturen, für die eine andere Skala gilt, müssen mit einem zusätzlichen refType temp_ (siehe dortige Definitionen) versehen werden
2023-11-15
Created
2024-12-06
Added XML example
2025-02-04: Adapted example to schema version 3.3.0 (replaced si:expandedUnc with si:measurementUncertaintyUnivariate)
2026-01-29
The use in dcc:influenceCondition is deprecated. Added example. Use a dcc:quantity inside dcc:influenceCondition instead for this refType.
<dcc:influenceConditions>
<dcc:influenceCondition refType="basic_ambient">
<dcc:name>
<dcc:content lang="de">Umgebungsbedingung</dcc:content>
<dcc:content lang="en">Environmental condition</dcc:content>
</dcc:name>
<dcc:description>
<dcc:content lang="de">Alle Werte wurden während der Kalibrierung überwacht.</dcc:content>
<dcc:content lang="en">All values were monitored during calibration.</dcc:content>
</dcc:description>
<dcc:data>
<dcc:quantity refType="basic_temperature">
<dcc:name>
<dcc:content lang="de">Umgebungsbedingung Temperatur</dcc:content>
<dcc:content lang="en">Environmental condition temperature</dcc:content>
</dcc:name>
<si:real>
<si:value>23.0</si:value>
<si:unit>\degreecelsius</si:unit>
<si:measurementUncertaintyUnivariate>
<si:expandedMU>
<si:valueExpandedMU>1</si:valueExpandedMU>
<si:coverageFactor>1.732</si:coverageFactor>
<si:coverageProbability>1</si:coverageProbability>
<si:distribution>uniform</si:distribution>
</si:expandedMU>
</si:measurementUncertaintyUnivariate>
</si:real>
</dcc:quantity>
</dcc:data>
</dcc:influenceCondition>
</dcc:influenceConditions>
<!-- example deprecated, use basic_temperature in dcc:quantity instead (see alternative XML example)-->
<dcc:influenceCondition refType="basic_temperature">
<dcc:name>
<dcc:content lang="de">Temperatur</dcc:content>
<dcc:content lang="en">temperature</dcc:content>
</dcc:name>
<dcc:data>
<dcc:quantity refType="basic_mean">
<si:real>
<si:value>294.05</si:value>
<si:unit>\kelvin</si:unit>
<si:measurementUncertaintyUnivariate>
<si:expandedMU>
<si:valueExpandedMU>0.50</si:valueExpandedMU >
<si:coverageFactor>2</si:coverageFactor>
<si:coverageProbability>0.95</si:coverageProbability>
</si:expandedMU>
</si:measurementUncertaintyUnivariate>
</si:real>
</dcc:quantity>
</dcc:data>
</dcc:influenceCondition>