Indicated values at zero load before load is applied.
Anzeigewerte, die im unbelasteten Zustand vor einer Belastung aufgezeichnet wurden.
2025-08-06 created
2026-04-15: added example for torque, added metadata for force example
ISO 376:2011 table 1
<dcc:quantity refType="torque_indicationAtZeroLoad">
<dcc:name>
<dcc:content lang="de">Nullanzeige</dcc:content>
<dcc:content lang="en">Indication at zero load</dcc:content>
</dcc:name>
<si:realListXMLList>
<si:valueXMLList>0.000217 0.000258 0.000273 0.000093 0.000240</si:valueXMLList>
<si:unitXMLList>\milli\volt\per\volt</si:unitXMLList>
</si:realListXMLList>
</dcc:quantity>
...
<dcc:measurementMetaData>
<dcc:metaData>
<dcc:data>
<dcc:quantity refType="torque_mountingPosition">
<dcc:name>
<dcc:content lang="de">Einbaustellung</dcc:content>
<dcc:content lang="en">Mounting position</dcc:content>
</dcc:name>
<si:realListXMLList>
<si:valueXMLList>0 0 0 120 240</si:valueXMLList>
<si:unitXMLList>\degree</si:unitXMLList>
</si:realListXMLList>
</dcc:quantity>
</dcc:data>
</dcc:metaData>
</dcc:measurementMetaData>
<dcc:quantity refType="force_indicationAtZeroLoad">
<dcc:name>
<dcc:content lang="de">Nullanzeige i₀</dcc:content>
<dcc:content lang="en">Indication at zero load</dcc:content>
</dcc:name>
<si:realListXMLList>
<si:valueXMLList>0.000000 0.000000 0.000000</si:valueXMLList>
<si:unitXMLList>\milli\volt\per\volt</si:unitXMLList>
</si:realListXMLList>
</dcc:quantity>
...
<dcc:measurementMetaData>
<dcc:metaData>
<dcc:data>
<dcc:quantity refType="torque_mountingPosition">
<dcc:name>
<dcc:content lang="de">Einbaustellung</dcc:content>
<dcc:content lang="en">Mounting position</dcc:content>
</dcc:name>
<si:realListXMLList>
<si:valueXMLList>0 120 240</si:valueXMLList>
<si:unitXMLList>\degree</si:unitXMLList>
</si:realListXMLList>
</dcc:quantity>
</dcc:data>
</dcc:metaData>
</dcc:measurementMetaData>