Procedure that was used to convert indications to measured values
Methode nach der die Konvertierung von Anzeigen in Messwerte erfolgte
//dcc:usedMethod[@refType="basic_conversionProcedure"]
Accounts for the conversion of indiations of the calibration object
Betrifft die Konvertierung der Anzeigen des Kalibriergegenstandes
2024-09-27: Created
2025-02-03: Adapted wording for indications/masured Values/calibration results to correct terms according to VIM.
DKD-R 5-1
<dcc:usedMethod refType="basic_conversionProcedure">
<dcc:name>
<dcc:content lang="en">Conversion procedure</dcc:content>
</dcc:name>
<dcc:description>
<dcc:content lang="en">Temperature values assigned to the measured resistance values according to DIN EN 60751:2022 are represented and compared to the reference values.</dcc:content>
</dcc:description>
<dcc:norm>DIN EN 60751:2022</dcc:norm>
</dcc:usedMethod>
<dcc:usedMethod refType="basic_conversionProcedure">
<dcc:name>
<dcc:content lang="de">Umrechnungsvorschrift</dcc:content>
</dcc:name>
<dcc:description>
<dcc:content lang="de">Aus den gemessenen Widerstandswerten wurden die entsprechenden Temperaturwerte gemäß DIN EN 60751:2022 ermittelt und den Temperatur-Referenzwerten gegenübergestellt.</dcc:content>
</dcc:description>
<dcc:norm>DIN EN 60751:2022</dcc:norm>
</dcc:usedMethod>