{"tema_id":"182","string":"relativeZeroError","created":"2024-06-05 10:18:35","code":null,"modified":"2024-07-11 05:10:53","notes":[{"@type":"Definition note","@lang":"de","@value":"Die relative Abweichung von Nullpunkten, die vor und nach jeder Messreihe aufgezeichnet werden und auf den Anzeigewert der maximalen Kalibrierkraft bezogen ist. Die relative Nullpunktabweichung ist in Prozent angegeben. "},{"@type":"Definition note","@lang":"en-GB","@value":"The relative deviation of zero readings recorded before and after each series of tests related to the indicated value at maximum calibration force. The relative zero error is provided in percent. "},{"@type":"Scope note","@lang":"en-GB","@value":"The zero reading shall be taken approximately 30 s after the force has been completely removed. "},{"@type":"Scope note","@lang":"de","@value":"Die Nullpunktablesung muss etwa 30 Sekunden nach der vollst\u00e4ndigen Entlastung vorgenommen werden. "},{"@type":"Historical note","@lang":"en","@value":"2025-09-03 created "},{"@type":"Historical note","@lang":"de","@value":"2025-09-03 erstellt "},{"@type":"Historical note","@lang":"en-US","@value":"2026-03-18: example updated "},{"@type":"Bibliographic note","@lang":"en","@value":"ISO 376: 2011 Chapter 7.5.3 "},{"@type":"Bibliographic note","@lang":"de","@value":"ISO 376: 2011 Kapitel 7.5.3 "},{"@type":"Example note","@lang":"en","@value":"\n&lt;dcc:result refType=\"force_relativeZeroError\"&gt;\n    &lt;dcc:name&gt;\n        &lt;dcc:content lang=\"de\"&gt;Relative Restanzeigen bezogen auf die H\u00f6chstkraft in %&lt;\/dcc:content&gt;\n        &lt;dcc:content lang=\"en\"&gt;Relative zero shift after preloading related to the maximum force in %&lt;\/dcc:content&gt;\n    &lt;\/dcc:name&gt;\n    &lt;dcc:description&gt;\n        &lt;dcc:content lang=\"de\"&gt;Kraftrichtung: Druckkraft&lt;\/dcc:content&gt;\n        &lt;dcc:content lang=\"en\"&gt;Force application: compressive&lt;\/dcc:content&gt;\n    &lt;\/dcc:description&gt;\n    &lt;dcc:data&gt;\n        &lt;!-- dcc.xsd v3.4.0 and later: use the tableDimension attribute to indicate that the following list represents a table in long format:\n        &lt;dcc:list tableDimension=\"1\" refType=\"force_afterPreloadings\"&gt; --&gt;\n        &lt;dcc:list refType=\"force_afterPreloadings\"&gt;\n            &lt;dcc:name&gt;\n                &lt;dcc:content lang=\"de\"&gt;Nach Vorbelastungen&lt;\/dcc:content&gt;\n                &lt;dcc:content lang=\"en\"&gt;After preloadings&lt;\/dcc:content&gt;\n            &lt;\/dcc:name&gt;\n            &lt;dcc:quantity&gt;\n                &lt;dcc:name&gt;\n                    &lt;dcc:content lang=\"de\"&gt;Relative Nullpunktabweichung f\u2080&lt;\/dcc:content&gt;\n                    &lt;dcc:content lang=\"en\"&gt;Relative zero error f\u2080&lt;\/dcc:content&gt;\n                &lt;\/dcc:name&gt;\n                &lt;si:realListXMLList&gt;\n                    &lt;si:valueXMLList&gt;0.001 0.002 0.003&lt;\/si:valueXMLList&gt;\n                    &lt;si:unitXMLList&gt;\\percent&lt;\/si:unitXMLList&gt;\n                &lt;\/si:realListXMLList&gt;\n            &lt;\/dcc:quantity&gt;\n            &lt;dcc:measurementMetaData&gt;\n                &lt;dcc:metaData&gt;\n                    &lt;dcc:data&gt;\n                        &lt;!-- dcc.xsd v3.4.0 and later: use the index attribute to indicate that the following quantity represents an index column of the table:\n                        &lt;dcc:quantity index=\"0\" refType=\"force_mountingPosition\"&gt; --&gt;\n                        &lt;dcc:quantity refType=\"force_mountingPosition\"&gt;\n                            &lt;dcc:name&gt;\n                                &lt;dcc:content lang=\"de\"&gt;Einbaustellung&lt;\/dcc:content&gt;\n                                &lt;dcc:content lang=\"en\"&gt;Mounting position&lt;\/dcc:content&gt;\n                            &lt;\/dcc:name&gt;\n                            &lt;si:realListXMLList&gt;\n                                &lt;si:valueXMLList&gt;0 120 240&lt;\/si:valueXMLList&gt;\n                                &lt;si:unitXMLList&gt;\\degree&lt;\/si:unitXMLList&gt;\n                            &lt;\/si:realListXMLList&gt;\n                        &lt;\/dcc:quantity&gt;\n                    &lt;\/dcc:data&gt;\n                &lt;\/dcc:metaData&gt;\n                &lt;dcc:metaData refId=\"functionalTest\"&gt;\n                    &lt;dcc:validXMLList&gt;true&lt;\/dcc:validXMLList&gt;\n                &lt;\/dcc:metaData&gt;\n            &lt;\/dcc:measurementMetaData&gt;\n        &lt;\/dcc:list&gt;\n        &lt;!-- dcc.xsd v3.4.0 and later: use the tableDimension attribute to indicate that the following list represents a table in long format:\n        &lt;dcc:list tableDimension=\"1\" refType=\"force_afterMeasurementSeries\"&gt; --&gt;\n        &lt;dcc:list refType=\"force_afterMeasurementSeries\"&gt;\n            &lt;dcc:name&gt;\n                &lt;dcc:content lang=\"de\"&gt;Nach Messreihen&lt;\/dcc:content&gt;\n                &lt;dcc:content lang=\"en\"&gt;After measurement series&lt;\/dcc:content&gt;\n            &lt;\/dcc:name&gt;\n            &lt;dcc:quantity&gt;\n                &lt;dcc:name&gt;\n                    &lt;dcc:content lang=\"de\"&gt;Relative Nullpunktabweichung f\u2080&lt;\/dcc:content&gt;\n                    &lt;dcc:content lang=\"en\"&gt;Relative zero error f\u2080&lt;\/dcc:content&gt;\n                &lt;\/dcc:name&gt;\n                &lt;si:realListXMLList&gt;\n                    &lt;si:valueXMLList&gt;0.006 0.007 0.008 0.009&lt;\/si:valueXMLList&gt;\n                    &lt;si:unitXMLList&gt;\\percent&lt;\/si:unitXMLList&gt;\n                &lt;\/si:realListXMLList&gt;\n            &lt;\/dcc:quantity&gt;\n            &lt;dcc:measurementMetaData&gt;\n                &lt;dcc:metaData&gt;\n                    &lt;dcc:data&gt;\n                        &lt;!-- dcc.xsd v3.4.0 and later: use the index attribute to indicate that the following quantity represents an index column of the table:\n                        &lt;dcc:quantity index=\"0\" refType=\"force_measurementSeries force_referenceList\"&gt; --&gt;\n                        &lt;dcc:quantity refType=\"force_measurementSeries force_referenceList\"&gt;\n                            &lt;dcc:name&gt;\n                                &lt;dcc:content lang=\"de\"&gt;Messreihe&lt;\/dcc:content&gt;\n                                &lt;dcc:content lang=\"en\"&gt;Measuring series&lt;\/dcc:content&gt;\n                            &lt;\/dcc:name&gt;\n                            &lt;dcc:noQuantity&gt;\n                                &lt;dcc:content refId=\"measurementSeries1Compressive\"&gt;X\u2081&lt;\/dcc:content&gt;\n                                &lt;dcc:content refId=\"measurementSeries2Compressive\"&gt;X\u2082&lt;\/dcc:content&gt;\n                                &lt;dcc:content refId=\"measurementSeries4Compressive\"&gt;X\u2084'&lt;\/dcc:content&gt;\n                                &lt;dcc:content refId=\"measurementSeries6Compressive\"&gt;X\u2086'&lt;\/dcc:content&gt;\n                            &lt;\/dcc:noQuantity&gt;\n                        &lt;\/dcc:quantity&gt;\n                    &lt;\/dcc:data&gt;\n                &lt;\/dcc:metaData&gt;\n                &lt;dcc:metaData refId=\"functionalTest\"&gt;\n                    &lt;dcc:validXMLList&gt;true&lt;\/dcc:validXMLList&gt;\n                &lt;\/dcc:metaData&gt;\n            &lt;\/dcc:measurementMetaData&gt;\n        &lt;\/dcc:list&gt;\n    &lt;\/dcc:data&gt;\n&lt;\/dcc:result&gt;\n "}]}